| Sign In | Join Free | My entremaqueros.com |
|
Brand Name : GE
Model Number : M1025230
MOQ : 1pcs
Payment Terms : T/T(wire transfer), Western Union, Paypal
Delivery Time : 3-5 working days
Packaging Details : Carton
Product Name : GE Datex Ohmeda E-COP Module M1025230 For Monitoring The State Of The Heart
Brand : GE
Part No : M1025230
Dimensions (H x W x D) : 11.2 x 3.7 x 18.7 cm (4.4 x 1.5 x 7.4 in)
Weight : 0.35 kg (0.77 lb)
Temperature : 10 to 40°C (50 to 104°F)
Relative humidity : 10 to 90% non-condensing
MOQ : One piece
Warranty : 90 days
GE Datex Ohmeda E-COP Module M1025230 For Monitoring The State Of The Heart
The E-COP module integrates the measurement of cardiac output (CO), right ventricular ejection fraction (REF) and invasive arterial pressure (P) by thermodilution on the CARESCAPE™ monitor screen. The E-COPSv module allows additional measurement of combined venous oxygen saturation (SvO2) and central venous oxygen saturation (ScvO2). The module complies with IEC 60601-1 3rd edition.
Features of GE E-COP Module
• The module contains direct keys to the C.O. to start. Measurement and zeroing of invasive pressure channels
• A unique set of automatic CO measurement types
• Up to six COs can edit the average CO measurement.
• Hemodynamic and oxygenation calculation screen view
• Provides the following hemodynamic calculations with indexed values: CI, SV, SI, SVR, SVRI, PVR, PVRI, LVSW, LVSWI, LCW, LCWI, RCW, RCWI, RVEDV, RVEDVI, RVSW, RVSWI+RVESV and RVESVI
• Provides the following oxygenation calculations with indexed values: CaO2, CvO2, C(a-v)O2, VO2, VO2I, RQ, ATMP, VO2 calculation, VO2I calculation, O2ER, DO2, DO2I, PaO2, AaDO2, CcO2, Qs /Qt, CO calculation. and CI calculations.
• Edit calculation input data
• Trend calculation and printing
• Module insertion/removal without interrupting other monitoring. • SvO2, ScvO2, CO and REF measurements are provided by Edwards Lifesciences Corp. compatible catheters.
Specification of GE E-COP Module
| Dimensions (H x W x D) | 11.2 x 3.7 x 18.7 cm (4.4 x 1.5 x 7.4 in) |
| Weight | 0.35 kg (0.77 lb) |
| Operating conditions | |
| Temperature | 10 to 40°C (50 to 104°F) |
| Relative humidity | 10 to 90% non-condensing |
| Storage conditions | |
| Temperature | -25 to 60°C (-13 to 140°F) |
| Relative humidity | 10 to 90% non-condensing |
| Waveform display | |
| Range | -30 to 300 mmHg |
| Scales | Adjustable in 10 mmHg increments |
| Numerical display | |
| Resolution | 1 mmHg |
| Alarms | Adjustable high and low alarm limits for systolic, diastolic and mean pressures |
| Pulse rate | |
| Measurement range | 30 to 250 bpm |
| Display resolution | 1 bpm |
| Accuracy | ±5% or ± 5 bpm |
| Invasive blood pressure | |
| Measurement range | -30 to 320 mmHg |
| Measurement accuracy | ±4% or ±4 mmHg |
| Transducer sensitivity | 5 µV/V/mmHg |
| Pressure filter | 0 to 22 Hz (-3 dB) adjustable upper limit 4 to 22 Hz |